The Model 7600 is Ivy Biomedical System’s 5th generation cardiac trigger monitor used in applications requiring precision ECG R-wave synchronization for gating host systems on and off. The Model 7600 incorporates an intuitive, menu-driven, touch-screen user interface, simplifying operation for clinicians and technicians.
- Intended use is primarily on patients in applications requiring precision R-wave synchronization
- Incorporates a simple, easy to use touch interface, the 7600 displays two simultaneous ECG vectors along with the patient's heart rate
- Manufactured by Ivy
Features
- Offers a compact 6″ diagonal color TFT display
- ECG simulator
- Heart rate limit alarms
- RS-232 communication interface
Specifications
ECG:
- Lead Selection
- Trigger Lead: LI, LII, LIII and Auto
- Second Lead: LI, LII, and LIII
Patient Cable
- 4 lead patient cable with 6 Pin AAMI standard connector.
- Isolated from ground related circuits by > 4 kV rms. 5.5 kV peak
Electrode Impedance Measurement
- Technique: 10Hz ac signal < 10 uA rms
- Range: 200 kOper lead
- Accuracy: ±3% ±1kO
- Recommended Electrode: 10% Chloride sponge type (590436)
Cardiotach
- Range: 10-350 BPM (Pediatric/Neonate); 10-300 BPM (Adult)
- Accuracy: ±1% ±1 BPM
- Resolution: 1BPM
- Sensitivity: 300 μV peak
- Tall T Wave Rejection: Rejects T waves = 1.2 * R-wave
Pacer Pulse Rejection
- Width: 0.1-2 ms at ±2 to ±700m V
Alarms
- High Rate: 15 to 250 BPM in 5 BPM increments
- Low Rate: 10 to 245 BPM in 5 BPM increments
- Asystole: R to R interval >6 seconds
Optional Recorder:
- Wiring Method: Direct Thermal
Synchronized Output (Trigger)
- Test input signal at ECG Leads
- Conditions: 1/2 sine wave, 60 ms width, 1mV amplitude, 1 pulse, second
- Output Trigger Delay: <10ms
- R to R Trigger Accuracy: ±75 μs typical@ 1 mV input
- Pulse Width: 100ms
- Pulse Amplitude: 0 to +5V
- Output Impedence: <100 O
Item Numbers
7600 Ivy Model 7600 Cardiac Trigger Monitor