Ludlum 14C with Energy Compensated Probe

Features a 5-range analog ratemeter and operates two detectors

Inquire About this Product

The Model 14C General Purpose Survey Meter features a 5-range analog ratemeter and operates two detectors; a built-in internal GM Detecter (0-2000 mR/hr) and an external GM or Scintillation Detector. The cast aluminum housing, with its separate battery compartment and accompanying metal handle, offer an industrial robustness and quality that promotes long-lasting protection and instrument life. The front panel controls include a rotary switch for selecting the five-decade range and instrument shut-off, an audio on/off switch, a fast/slow response switch, a high-voltage display button and a battery test button. A 39" straight type detector cable with "C" style connector is included.

Specifications

  • Application: Alpha, Beta and Gamma surveying; Energy compensated GM (used with x 1000 scale only)

  • Internal Detector: Range 0-20 mSv/hr (0-2000 mR/hr); Sensitivity 200 cpm per mR/hr; Compatible

  • External Detectors: Any GM or Scintillation operating on 900 VDC

  • Detector Channels:

    • External: x0.1 through x100 range settings

    • Internal GM: x1000 setting

  • Energy Response: within 15% of true value between 60 keV-3 MeV (internal detector only)

  • Linearity: reading within 10% of true value with detector connected

  • Meter Dial: 0-2 mR/hr, 0-5 kcpm, BAT TEST (others available)

  • Meter: 6.4 cm (2.5 in) arc, 1 mA analog type

  • Rotary Selector Switch: Switches between Off,

  • Ranges: x0.1, x1, x10, x100, x1000

  • Response Switch: Toggles between FAST (4 sec) or SLOW (22 sec) from 10% to 90% of final reading

  • Controls:

    • Reset Push Button: Zeroes meter

    • Battery Push Button: Checks battery status

    • Audio Switch: Audio on/off, built-in unimorph speaker, greater than 60 dB at 0.6 meters (2 ft)

  • High Voltage: 900 Vdc

  • Threshold: -40 mV to 10 mV

  • Temperature Range: -20 to 50o C (-4 to 122o F)

  • Relative Humidity: < 95%

  • Power: 2 each “D” cell batteries

  • Battery Life: Typically greater than 2000 hours

  • Construction: Cast and drawn aluminum with beige powder coat

  • Connector: Series “C” (others available)

  • Size: 6.5″ H x 3.5″ W x 8.5″ L

  • Weight: 3.5 lb (including batteries)

Item Numbers

051-4438 Ludlum 14C with Energy Compensated Probe

Related Items

051-196 Check Source Cs-137 1.0 µCi

051-197 Check Source With Stick On Holder

Cs-137 0.25 µCi

051-166 Check Source Holder