The Ludlum Model 3000 is a durable, lightweight instrument that is used with an external detector for alpha, beta, and gamma radiation survey. It features a large, easy-to-read LCD screen and audible alarms and is controlled using a simple five-button interface. The unit body is made of high-impact plastic, and splash-resistant construction allows the instrument to be used outdoors.

Three modes of operation are available – RATE, MAX, and COUNT – which can be selected by pressing the MODE button. Measurements can be collected in two sets of units (primary and secondary) for RATE and MAX modes in cps, cpm, Bq, dpm, mR/hr, or µSv/h units. The user can switch between two sets of chosen units by pressing the UNITS button. When enabled with the optional Lumic Data Logger Kit (see Options), data can be logged in any of the operational modes using the LOG button on the handle. Up to 1000 data points can be stored internally.

Instrument setup can be done either through the front-panel controls or via the Lumic Calibration Kit (see Options). The Model 3000 is shipped ready to use with batteries and calibration certificate.

Features

  • Large, Backlit, Easy-To-Read LCD Screen

  • Auto-Ranging

  • RATE, MAX, and COUNT Modes of Operation

  • Splash-Resistant Construction for Outdoor Use

  • All-Digital Calibration

  • USB Port

  • Lightweight Yet Ruggedly Built

  • Simple 5-Button Interface

Specifications

  • Compatible Detector:s GM, scintillation, or proportional

  • Display: 3 digit auto-ranging LCD with large 20 mm (0.8 in.) digits.

    • Units: (k)cps, (k)cpm, (k)Bq, (k)dpm, (µ)(m)R(/h), (µ)(m)Sv(/h).

    • Indicators: bar graph, MAX, ALARM, USB, audio, low battery.

  • Display Range

    • 0.0 cps to 99.9 kcps

    • 0.00 cpm to 999 kcpm

    • 0.00 Bq to 99.9 kBq

    • 0.00 dpm to 999 kdpm

    • 0.00 µR/h to 999 R/h

    • 0.00 µSv/h to 999 Sv/h

    • Minimum and maximum display can be set to limit display to calibrated range

  • Linearity Readings within 10% of true value

  • Backlight: Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-On (will reduce battery life)

  • Controls:

    • ON/OFF: Extra-long press to turn ON; Short press to acknowledge alarms; Extra-long press plus 3 seconds to turn OFF

    • UNITS: Short press changes the units between count rate, dose/exposure, or disintegration; Extra-long press to change count time in COUNT mode

    • AUDIO: Short press to adjust audio level

    • MODE: Short press alternates between available modes – RATE (displays count rate), MAX (captures peak rate), and COUNT (captures rate or dose readings for a user-selectable count time from 0 to 10 minutes)

    • LOG: Short press to log current display (requires Lumic Data Logger Kit), or initiate count i COUNT mode

  • Data Logging: Can store up to 1000 data points internally. Requires Lumic Data Logger Kit, see Options.

  • Response Time: User-selectable from 1 to 60 seconds, or auto-response rate FAST or SLOW

  • Alarms: Two count rate, exposure/dose, and count alarm setpoints adjustable over the display range

  • Overload Protection: High count rate saturation protection prevents false display of lower count rates. “OL” will appear on the display and an audio warning will be triggered.

  • Loss of Count Protection: After a user-defined time of no pulses from detector, typically 60 seconds, the unit will flash a zero reading, and an audio alarm will be triggered.

  • Overrange: If reading exceeds the predefined detector range, the instrument will flash a maximum reading and trigger an audio warning.

  • Dead Time Correction: Employs first and second order corrections for extended performance

  • High Voltage: 400 to 1500 Vdc

  • Threshold: -4 to -100 mVdc

  • Audio Approximately 4.5 kHz; Click audio greater than 65 dB at 0.6 m (2 ft); Alarm audio greater than 72 dB

  • Power: Four alkaline “AA” batteries

  • Battery Life: Approximately 750 hours of operation (as low as 100 hours with backlight configured for continuous-on), 16-hour low battery warning

  • Construction: High-impact plastic with water-resistant rubber seals and separate battery compartment

  • Temperature Range: -20 to 50 °C (-5 to 122 °F)

    • May be certified for operation from -40 to 65 °C (-40 to 150 °F)

  • Environmental Rating NEMA rating of 5 or IP rating of 53

  • Size (H x W x L) 16.5 x 11.4 x 21.6 cm (6.5 x 4.5 x 8.5 in.)

  • Weight 0.7 kg (1.6 lb) (excluding detectors and clips)

Item Numbers

48-4035 Model 3000 Digital Survey Meter